HHD-01


SPECIFICATION
System:-
Microprocessor controlled heated humidifier.
Display:-
Digital display of set temperature and gas temperature in chamber.
Control:-
Maintains chamber gas temperature within ±0.3 C of set value.
Safety:-
High Temperature Cutoff, Overheat Protection.
Other Features:-
: Easily mountable, slide-on, transparent, autoclavable water chamber for passover humidification. Ease of compatibility with most ventilators and anesthesia machines.
Power:-
230V A.C. 50Hz.
Tested as per following:-
As per CISPR 11:2019, IEC 61000-4-6:2013, IEC 61000-4-4:2012, IEC 61000-4-5:2017, IEC 61000-4-8:2009 & IEC 61000-4-2:2008
IEC 60601-1:2005+A1:2012+A2:2020
IEC 60601-1:2005+AMD1:2012+AMD2:2020
IEC 60601-1-8:2006+A1:2012+A2:2020
IEC61000-3-2 Edition 5.0 2020-07
HHD-01
HD-PLUS

SPECIFICATION
System:-
Microprocessor Controlled Servo
Heated Humidifier.
State-of-the-art Heated Wire
Technology for controlled
Humidification at Distal End.
Display:-
Digital Display of Gas Temperature at Distal End.
Control:-
Maintains Gas Temperature within ±0.3 C of set value.
Safety:-
High Temperature Cutoff, Overheat Protection.
Other Features:-
Reusable Heated Wire for Economy in Usage.
Easily Mountable, Side-on,
Transparent Autoclavable Water
Chamber for Passover Humidification.
Ease of Compatibility with Most Ventilators and Anaesthesia Machines.
Power:-
230V A.C. 50Hz.
ACCESSORIES:-
Humidifier Chamber, Sensor wire, Mains Chord.
HD-PLUS
HD-PRO

SPECIFICATION
System Specifications:-
- System: Microprocessor Controlled Servo Heated Humidifier.
- Technology: State-of-the-art heated wire technology for controlled humidification at distal End.
- Display: Digital display of gas temperature at chamber as well as patient end
- Control: Maintains gas temperature within ±0.3˚C of default set value.
Technical Specifications:-
Alarms:
- Chamber Temperature High/Low (Audio and Visual)
- Airway Temperature High/Low (Audio and Visual)
- Temperature Probe Open (Audio and Visual)
- Inspiratory heater and Expiratory heater ON/OFF (Visual)
Power: 230V A.C, 50Hz.
Supply Current: 1 A max at 230V
Power of Heating Tray: 150w
Heater Wire capacity: 22 VAC
Heater Plate Thermal Cut-off: 120˚C ±5 ˚C
Display Temperature Range: 15 ˚C to 70 ˚C
Key Features:-
- Modes: Invasive and Non-Invasive
- Reusable Heated Wire for Economy in Usage
- Side-on clamp to mount easily
- Transparent autoclavable water chamber for pass-over humidification
- Ease of Compatibility with Most CPAP, Bubble CPAP, HFNC CPAP, Ventilators and Anaesthesia Machines
Description | Invasive | Non-Invasive | Heater Wire Status |
Chamber Set Point | 37˚C | 31˚C | ON |
Airway Set Point | 40˚C | 34˚C | ON |
Chamber Set Point | Less than 66˚C | Less than 66˚C | OFF |
Airway Set Point | 37˚C | 31˚C | OFF |