INSPIRATION (NPRW-97)
Key features
- On board temperature calibration facility with feather touch keys
- Entire warmer height updown facility with motorized remote control optional
- Baby tray up down trendelenburg facility optional with motorized remote control
- Swiveling 3 drawers
- Blender (optional)
Components
Micro Computer Temperature Controller
- PID (proportionate integrated derivative) based micro processor technology
- Dual monitoring of Central and Peripheral temperature with early
- warning of Cold Stress onset
- User friendly on-board temperature calibration with feather touch keys
- User friendly feather touch control
- Key-pad lock facility
- Safety alarms along with high temperature cutoff
- Watch Dog facility
Heater Box
- Ceramic Infra Red Heater
- Swivelling source box to take X-ray
- Stainless steel parabolic reflector with heater grill
- LED Examination Lamp
Baby Bassinet
- Easy access of an Infant from all sides due to clear acrylic collapsible side panels
- Provision for positioning X-Ray cassette
- Thick cushion mattress
Optional Attachment
- Masimo SPO2 (product code: NPRW-97)
- Inline Weighing
- T-Piece Resuscitator
- Neonatal Suction
- Heater Source remains Focussed on Baby even in Swivel condition
Skin/Air/Manual
Displays
- Skin temperature
- Air temperature
- Peripheral temperature (optional)
- Heater output (%)
- Set temperature
- Control mode
- APGAR Timer
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Auto Mode
Safety alarm messages on LCD
Skin Mode
- “Skin Temp. High” (>0.5°C of set temp.)
- “Skin Temp. Low” (<0.5°C of set temp.)
- “Skin Temp. Over Range” (>38°C)
- “Skin Temp. Over Range” (<34°C)
- “Skin Probe Fail”
Air Mode
- “Air Temp. High” (>1.5°C of set temp.)
- “Air Temp. Low” (<3.0°C of set temp.)
- “Air Temp. Over Range” (>39°C)
- “Air Probe Fail”
Safety high temp. cutoff at
- Skin temp. >38°C
- Air temp. >39°C
- Power Fail
- Heater Fail (optional)
Cold Stress Alarm (optional)
Peripheral Temp. 1.5°C < Skin Temp.
Power Consumption : 800 Watt
Tests
Tested as per following IEC Standards :
IEC 601-1:1998 + A1:1991 + A2:1995
IEC 601-2-21:1994
IEC 60601-1-2, 2001, 2nd Edition
Amendment A1:1996 to IEC 60601-2-21:1994
INSPIRATION (NPRW-97)
INNOVA (NPRHWINV-01)- FIXED BASSINET + DOUBLESURFACE PHOTOTHERAPY
Components
Micro Computer Temperature Controller
- PID (proportionate integrated derivative) based micro processor technology assures accurate maintenance of set temperature with internal accuracy of 0.01°C
- User friendly on-board temperature calibration with feather touch keys
- User friendly feather touch control
- Key-pad lock facility to avoid accidental changes of any set parameters
- Safety alarm with high temperature cutoff
- Watch Dog facility Auto-Resets the System during hangups due to power fluctuation
Heater Box
- Ceramic Infra Red Heater with higher power output and more durability than any tubular silicon quartz glass heaters
- Swivelling source box to take X-ray
- Temperature Resistant Design of heater box keeps exterior surface cool even after continuous use
- Stainless steel reflector and heater grill for complete cleaning of heater & reflector without any electrical shock hazards
LED Examination Lamp
- Appropriate illumination to assess infant colour during procedures
Baby Bassinet (Stainless Steel)
- Easy access of an Infant from all sides due to clear acrylic collapsible side panels
- Acrylic Baby Tray for infant’s head Up / Down facility
- Provision for positioning X-Ray cassette
- Thick cushion mattress
- Nylon based moulded side pillars to fix the drop down sides
Essential Attachments
- Instrument Tray, Pole for Infusion Pump & IV Stand
Available Options
- Fabrication in Mild Steel or Stainless Steel
- Air/Pre-warm, Skin/Manual or Skin/Air/Manual (LCD) Controller
Temperature Controllers
Skin / Air / Manual (LCD)
Displays
- 1” bright skin temp. display
- 1” bright air temp. display
- Numeric % of heater output display
- Set temp. display on LCD
- Control mode display on LCD
- APGAR Timer display on LCD
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Auto Mode
Safety alarm messages on LCD
Skin Mode
- “Skin Temp. High” (>0.5°C of set temp.)
- “Skin Temp. Low” (<0.5°C of set temp.)
- “Skin Temp. Over Range” (>38°C)
- “Skin Temp. Over Range” (<34°C)
- “Skin Probe Fail”
Air Mode
- “Air Temp. High” (>1.5°C of set temp.)
- “Air Temp. Low” (<3.0°C of set temp.)
- “Air Temp. Over Range” (>39°C)
- “Air Probe Fail”
Safety high temp. cutoff at
- Skin temp. >38°C
- Air temp. >39°C
- Power Fail
- Heater Fail (optional)
Skin / Manual
Display
- 1” bright skin temp. display
- Set temp. display
- Numeric % of heater output display
- Selection of Skin & Manual modes
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Skin Mode
Safety alarm messages on LCD
- “Temp. High” (<0.5°C of set temp.)
- “Temp. Low” (>0.5°C of set temp.)
- “Temp. Over Range” (>38°C) /
- “Temp. Under Range” (<34°C)
- (Actual temp. starts blinking)
- “Probe Fail”
- Power failure
Safety high temp. cutoff at 38°C
For air control low temp. limit (<34°C) can be programmed.
INNOVA (NPRHWINV-01)- FIXED BASSINET + DOUBLESURFACE PHOTOTHERAPY
INNOVA (NPRHWINV-02)- FIXED BASSINET + UNDERSURFACE PHOTOTHERAPY
Components
Micro Computer Temperature Controller
- PID (proportionate integrated derivative) based micro processor technology assures accurate maintenance of set temperature with internal accuracy of 0.01°C
- User friendly on-board temperature calibration with feather touch keys
- User friendly feather touch control
- Key-pad lock facility to avoid accidental changes of any set parameters
- Safety alarm with high temperature cutoff
- Watch Dog facility Auto-Resets the System during hangups due to power fluctuation
Heater Box
- Ceramic Infra Red Heater with higher power output and more durability than any tubular silicon quartz glass heaters
- Swivelling source box to take X-ray
- Temperature Resistant Design of heater box keeps exterior surface cool even after continuous use
- Stainless steel reflector and heater grill for complete cleaning of heater & reflector without any electrical shock hazards
LED Examination Lamp
- Appropriate illumination to assess infant colour during procedures
Baby Bassinet (Stainless Steel)
- Easy access of an Infant from all sides due to clear acrylic collapsible side panels
- Acrylic Baby Tray for infant’s head Up / Down facility
- Provision for positioning X-Ray cassette
- Thick cushion mattress
- Nylon based moulded side pillars to fix the drop down sides
Essential Attachments
- Instrument Tray, Pole for Infusion Pump & IV Stand
Available Options
- Fabrication in Mild Steel or Stainless Steel
- Air/Pre-warm, Skin/Manual or Skin/Air/Manual (LCD) Controller
Temperature Controllers
Skin / Air / Manual (LCD)
Displays
- 1” bright skin temp. display
- 1” bright air temp. display
- Numeric % of heater output display
- Set temp. display on LCD
- Control mode display on LCD
- APGAR Timer display on LCD
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Auto Mode
Safety alarm messages on LCD
Skin Mode
- “Skin Temp. High” (>0.5°C of set temp.)
- “Skin Temp. Low” (<0.5°C of set temp.)
- “Skin Temp. Over Range” (>38°C)
- “Skin Temp. Over Range” (<34°C)
- “Skin Probe Fail”
Air Mode
- “Air Temp. High” (>1.5°C of set temp.)
- “Air Temp. Low” (<3.0°C of set temp.)
- “Air Temp. Over Range” (>39°C)
- “Air Probe Fail”
Safety high temp. cutoff at
- Skin temp. >38°C
- Air temp. >39°C
- Power Fail
- Heater Fail (optional)
Skin / Manual
Display
- 1” bright skin temp. display
- Set temp. display
- Numeric % of heater output display
- Selection of Skin & Manual modes
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Skin Mode
Safety alarm messages on LCD
- “Temp. High” (<0.5°C of set temp.)
- “Temp. Low” (>0.5°C of set temp.)
- “Temp. Over Range” (>38°C) /
- “Temp. Under Range” (<34°C)
- (Actual temp. starts blinking)
- “Probe Fail”
- Power failure
Safety high temp. cutoff at 38°C
For air control low temp. limit (<34°C) can be programmed.
INNOVA (NPRHWINV-02)- FIXED BASSINET + UNDERSURFACE PHOTOTHERAPY
INNOVA (NPRHWINV-03)- FIXED BASSINET + DRAWERS
Components
Micro Computer Temperature Controller
- PID (proportionate integrated derivative) based micro processor technology assures accurate maintenance of set temperature with internal accuracy of 0.01°C
- User friendly on-board temperature calibration with feather touch keys
- User friendly feather touch control
- Key-pad lock facility to avoid accidental changes of any set parameters
- Safety alarm with high temperature cutoff
- Watch Dog facility Auto-Resets the System during hangups due to power fluctuation
Heater Box
- Ceramic Infra Red Heater with higher power output and more durability than any tubular silicon quartz glass heaters
- Swivelling source box to take X-ray
- Temperature Resistant Design of heater box keeps exterior surface cool even after continuous use
- Stainless steel reflector and heater grill for complete cleaning of heater & reflector without any electrical shock hazards
LED Examination Lamp
- Appropriate illumination to assess infant colour during procedures
Baby Bassinet (Stainless Steel)
- Easy access of an Infant from all sides due to clear acrylic collapsible side panels
- Acrylic Baby Tray for infant’s head Up / Down facility
- Provision for positioning X-Ray cassette
- Thick cushion mattress
- Nylon based moulded side pillars to fix the drop down sides
Essential Attachments
- Instrument Tray, Pole for Infusion Pump & IV Stand
Available Options
- Fabrication in Mild Steel or Stainless Steel
- Air/Pre-warm, Skin/Manual or Skin/Air/Manual (LCD) Controller
Temperature Controllers
Skin / Air / Manual (LCD)
Displays
- 1” bright skin temp. display
- 1” bright air temp. display
- Numeric % of heater output display
- Set temp. display on LCD
- Control mode display on LCD
- APGAR Timer display on LCD
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Auto Mode
Safety alarm messages on LCD
Skin Mode
- “Skin Temp. High” (>0.5°C of set temp.)
- “Skin Temp. Low” (<0.5°C of set temp.)
- “Skin Temp. Over Range” (>38°C)
- “Skin Temp. Over Range” (<34°C)
- “Skin Probe Fail”
Air Mode
- “Air Temp. High” (>1.5°C of set temp.)
- “Air Temp. Low” (<3.0°C of set temp.)
- “Air Temp. Over Range” (>39°C)
- “Air Probe Fail”
Safety high temp. cutoff at
- Skin temp. >38°C
- Air temp. >39°C
- Power Fail
- Heater Fail (optional)
Skin / Manual
Display
- 1” bright skin temp. display
- Set temp. display
- Numeric % of heater output display
- Selection of Skin & Manual modes
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Skin Mode
Safety alarm messages on LCD
- “Temp. High” (<0.5°C of set temp.)
- “Temp. Low” (>0.5°C of set temp.)
- “Temp. Over Range” (>38°C) /
- “Temp. Under Range” (<34°C)
- (Actual temp. starts blinking)
- “Probe Fail”
- Power failure
Safety high temp. cutoff at 38°C
For air control low temp. limit (<34°C) can be programmed.
INNOVA (NPRHWINV-03)- FIXED BASSINET + DRAWERS
INNOVA (NPRHWINV-04)- FIXED BASSINET
Components
Micro Computer Temperature Controller
- PID (proportionate integrated derivative) based micro processor technology assures accurate maintenance of set temperature with internal accuracy of 0.01°C
- User friendly on-board temperature calibration with feather touch keys
- User friendly feather touch control
- Key-pad lock facility to avoid accidental changes of any set parameters
- Safety alarm with high temperature cutoff
- Watch Dog facility Auto-Resets the System during hangups due to power fluctuation
Heater Box
- Ceramic Infra Red Heater with higher power output and more durability than any tubular silicon quartz glass heaters
- Swivelling source box to take X-ray
- Temperature Resistant Design of heater box keeps exterior surface cool even after continuous use
- Stainless steel reflector and heater grill for complete cleaning of heater & reflector without any electrical shock hazards
LED Examination Lamp
- Appropriate illumination to assess infant colour during procedures
Baby Bassinet (Stainless Steel)
- Easy access of an Infant from all sides due to clear acrylic collapsible side panels
- Acrylic Baby Tray for infant’s head Up / Down facility
- Provision for positioning X-Ray cassette
- Thick cushion mattress
- Nylon based moulded side pillars to fix the drop down sides
Essential Attachments
- Instrument Tray, Pole for Infusion Pump & IV Stand
Available Options
- Fabrication in Mild Steel or Stainless Steel
- Air/Pre-warm, Skin/Manual or Skin/Air/Manual (LCD) Controller
Temperature Controllers
Skin / Air / Manual (LCD)
Displays
- 1” bright skin temp. display
- 1” bright air temp. display
- Numeric % of heater output display
- Set temp. display on LCD
- Control mode display on LCD
- APGAR Timer display on LCD
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Auto Mode
Safety alarm messages on LCD
Skin Mode
- “Skin Temp. High” (>0.5°C of set temp.)
- “Skin Temp. Low” (<0.5°C of set temp.)
- “Skin Temp. Over Range” (>38°C)
- “Skin Temp. Over Range” (<34°C)
- “Skin Probe Fail”
Air Mode
- “Air Temp. High” (>1.5°C of set temp.)
- “Air Temp. Low” (<3.0°C of set temp.)
- “Air Temp. Over Range” (>39°C)
- “Air Probe Fail”
Safety high temp. cutoff at
- Skin temp. >38°C
- Air temp. >39°C
- Power Fail
- Heater Fail (optional)
Skin / Manual
Display
- 1” bright skin temp. display
- Set temp. display
- Numeric % of heater output display
- Selection of Skin & Manual modes
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Skin Mode
Safety alarm messages on LCD
- “Temp. High” (<0.5°C of set temp.)
- “Temp. Low” (>0.5°C of set temp.)
- “Temp. Over Range” (>38°C) /
- “Temp. Under Range” (<34°C)
- (Actual temp. starts blinking)
- “Probe Fail”
- Power failure
Safety high temp. cutoff at 38°C
For air control low temp. limit (<34°C) can be programmed.
INNOVA (NPRHWINV-04)- FIXED BASSINET
INNOVA (NPRHWLT-01) - FIXED BASSINET LITE
Components
Micro Computer Temperature Controller
- PID (proportionate integrated derivative) based micro processor technology assures accurate maintenance of set temperature with internal accuracy of 0.01°C
- User friendly on-board temperature calibration with feather touch keys
- User friendly feather touch control
- Key-pad lock facility to avoid accidental changes of any set parameters
- Safety alarm with high temperature cutoff
- Watch Dog facility Auto-Resets the System during hangups due to power fluctuation
Heater Box
- Ceramic Infra Red Heater with higher power output and more durability than any tubular silicon quartz glass heaters
- Swivelling source box to take X-ray
- Temperature Resistant Design of heater box keeps exterior surface cool even after continuous use
- Stainless steel reflector and heater grill for complete cleaning of heater & reflector without any electrical shock hazards
LED Examination Lamp
- Easy access of an Infant from all sides due to clear acrylic collapsible side panels
- Bed head Up / Down facility
- Provision for positioning X-Ray cassette
- Thick cushion mattress
- Basinet side cover easy to be open & close without difficulty
- Self locking single hand operation
- The side rails foldable to care the baby and transparent basinet tilting facility
Baby Bassinet (Stainless Steel)
- Easy access of an Infant from all sides due to clear acrylic collapsible side panels
- Acrylic Baby Tray for infant’s head Up / Down facility
- Provision for positioning X-Ray cassette
- Thick cushion mattress
- Nylon based moulded side pillars to fix the drop down sides
Essential Attachments
- Instrument Tray, Pole for Infusion Pump & IV Stand
- MCB (Optional)
- Manual Height adjustment (Optional)
Skin / Manual Temperature Controller
Displays
- 1” bright skin temp. display
- Set temp. display
- Numeric % of heater output display
- Selection of Skin & Manual modes
Manual Mode
- Time programmable
- Heater output adjustable
- Over temp. cutoff.
- Timer / APGAR Timer
- Audio Alarm mute facility
- Mute time programmable
Skin Mode
Safety alarm messages on LCD
- “Temp. High” (<0.5°C of set temp.)
- “Temp. Low” (>0.5°C of set temp.)
- “Temp. Over Range” (>38°C) / “Temp. Under Range” (<34°C) (Actual temp. starts blinking)
- “Probe Fail”
- Power failure
Safety high temp. cutoff at 38°C
- Skin Temp. – 28-37°C
- Display Temp. – 10-50°C
- APGAR Timer
- Working Temp.: 10-40°C
- Relative Humidity: 25-75%
- Atmospheric Pressure: 500 – 900 mmHg
- AC – 220 VAC – 50 Hz
- 4 Castor wheels, 2 lockable
Tests
Tested as per following IEC Standards :
IEC 601-1:1998 + A1:1991 + A2:1995
IEC 601-2-21:1994
IEC 60601-1-2, 2001, 2nd Edition
Amendment A1:1996 to IEC 60601-2-21:1994
INNOVA (NPRHWLT-01) - FIXED BASSINET LITE
INNOVA (NPRHWINV) - STAND
Components
Micro Computer Temperature Controller
- PID (proportionate integrated derivative) based micro processor technology assures accurate maintenance of set temperature with internal accuracy of 0.01°C
- User friendly on-board temperature calibration with feather touch keys
- User friendly feather touch control
- Key-pad lock facility to avoid accidental changes of any set parameters
- Safety alarm with high temperature cutoff
- Watch Dog facility Auto-Resets the System during hangups due to power fluctuation
Heater Box
- Ceramic Infra Red Heater with higher power output and more durability than any tubular silicon quartz glass heaters
- Swivelling source box to take X-ray
- Temperature Resistant Design of heater box keeps exterior surface cool even after continuous use
- Stainless steel reflector and heater grill for complete cleaning of heater & reflector without any electrical shock hazards
LED Examination Lamp
- Appropriate illumination to assess infant colour during procedures
Baby Bassinet (Stainless Steel)
- Easy access of an Infant from all sides due to clear acrylic collapsible side panels
- Acrylic Baby Tray for infant’s head Up / Down facility
- Provision for positioning X-Ray cassette
- Thick cushion mattress
- Nylon based moulded side pillars to fix the drop down sides
Essential Attachments
- Instrument Tray, Pole for Infusion Pump & IV Stand
Available Options
- Fabrication in Mild Steel or Stainless Steel
- Air/Pre-warm, Skin/Manual or Skin/Air/Manual (LCD) Controller
Temperature Controllers
Skin / Air / Manual (LCD)
Displays
- 1” bright skin temp. display
- 1” bright air temp. display
- Numeric % of heater output display
- Set temp. display on LCD
- Control mode display on LCD
- APGAR Timer display on LCD
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Auto Mode
Safety alarm messages on LCD
Skin Mode
- “Skin Temp. High” (>0.5°C of set temp.)
- “Skin Temp. Low” (<0.5°C of set temp.)
- “Skin Temp. Over Range” (>38°C)
- “Skin Temp. Over Range” (<34°C)
- “Skin Probe Fail”
Air Mode
- “Air Temp. High” (>1.5°C of set temp.)
- “Air Temp. Low” (<3.0°C of set temp.)
- “Air Temp. Over Range” (>39°C)
- “Air Probe Fail”
Safety high temp. cutoff at
- Skin temp. >38°C
- Air temp. >39°C
- Power Fail
- Heater Fail (optional)
Skin / Manual
Display
- 1” bright skin temp. display
- Set temp. display
- Numeric % of heater output display
- Selection of Skin & Manual modes
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Skin Mode
Safety alarm messages on LCD
- “Temp. High” (<0.5°C of set temp.)
- “Temp. Low” (>0.5°C of set temp.)
- “Temp. Over Range” (>38°C) /
- “Temp. Under Range” (<34°C)
- (Actual temp. starts blinking)
- “Probe Fail”
- Power failure
Safety high temp. cutoff at 38°C
For air control low temp. limit (<34°C) can be programmed.
INNOVA (NPRHWINV) - STAND
INSPIRATION (NPRW-97)
Key features
- On board temperature calibration facility with feather touch keys
- Entire warmer height updown facility with motorized remote control optional
- Baby tray up down trendelenburg facility optional with motorized remote control
- Swiveling 3 drawers
- Blender (optional)
Components
Micro Computer Temperature Controller
- PID (proportionate integrated derivative) based micro processor technology
- Dual monitoring of Central and Peripheral temperature with early
- warning of Cold Stress onset
- User friendly on-board temperature calibration with feather touch keys
- User friendly feather touch control
- Key-pad lock facility
- Safety alarms along with high temperature cutoff
- Watch Dog facility
Heater Box
- Ceramic Infra Red Heater
- Swivelling source box to take X-ray
- Stainless steel parabolic reflector with heater grill
- LED Examination Lamp
Baby Bassinet
- Easy access of an Infant from all sides due to clear acrylic collapsible side panels
- Provision for positioning X-Ray cassette
- Thick cushion mattress
Optional Attachment
- Masimo SPO2 (product code: NPRW-97)
- Inline Weighing
- T-Piece Resuscitator
- Neonatal Suction
- Heater Source remains Focussed on Baby even in Swivel condition
Skin/Air/Manual
Displays
- Skin temperature
- Air temperature
- Peripheral temperature (optional)
- Heater output (%)
- Set temperature
- Control mode
- APGAR Timer
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Auto Mode
Safety alarm messages on LCD
Skin Mode
- “Skin Temp. High” (>0.5°C of set temp.)
- “Skin Temp. Low” (<0.5°C of set temp.)
- “Skin Temp. Over Range” (>38°C)
- “Skin Temp. Over Range” (<34°C)
- “Skin Probe Fail”
Air Mode
- “Air Temp. High” (>1.5°C of set temp.)
- “Air Temp. Low” (<3.0°C of set temp.)
- “Air Temp. Over Range” (>39°C)
- “Air Probe Fail”
Safety high temp. cutoff at
- Skin temp. >38°C
- Air temp. >39°C
- Power Fail
- Heater Fail (optional)
Cold Stress Alarm (optional)
Peripheral Temp. 1.5°C < Skin Temp.
Power Consumption : 800 Watt
Tests
Tested as per following IEC Standards :
IEC 601-1:1998 + A1:1991 + A2:1995
IEC 601-2-21:1994
IEC 60601-1-2, 2001, 2nd Edition
Amendment A1:1996 to IEC 60601-2-21:1994
INSPIRATION (NPRW-97)
INNOVA (NPRHWINV-01)- FIXED BASSINET + DOUBLESURFACE PHOTOTHERAPY
Components
Micro Computer Temperature Controller
- PID (proportionate integrated derivative) based micro processor technology assures accurate maintenance of set temperature with internal accuracy of 0.01°C
- User friendly on-board temperature calibration with feather touch keys
- User friendly feather touch control
- Key-pad lock facility to avoid accidental changes of any set parameters
- Safety alarm with high temperature cutoff
- Watch Dog facility Auto-Resets the System during hangups due to power fluctuation
Heater Box
- Ceramic Infra Red Heater with higher power output and more durability than any tubular silicon quartz glass heaters
- Swivelling source box to take X-ray
- Temperature Resistant Design of heater box keeps exterior surface cool even after continuous use
- Stainless steel reflector and heater grill for complete cleaning of heater & reflector without any electrical shock hazards
LED Examination Lamp
- Appropriate illumination to assess infant colour during procedures
Baby Bassinet (Stainless Steel)
- Easy access of an Infant from all sides due to clear acrylic collapsible side panels
- Acrylic Baby Tray for infant’s head Up / Down facility
- Provision for positioning X-Ray cassette
- Thick cushion mattress
- Nylon based moulded side pillars to fix the drop down sides
Essential Attachments
- Instrument Tray, Pole for Infusion Pump & IV Stand
Available Options
- Fabrication in Mild Steel or Stainless Steel
- Air/Pre-warm, Skin/Manual or Skin/Air/Manual (LCD) Controller
Temperature Controllers
Skin / Air / Manual (LCD)
Displays
- 1” bright skin temp. display
- 1” bright air temp. display
- Numeric % of heater output display
- Set temp. display on LCD
- Control mode display on LCD
- APGAR Timer display on LCD
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Auto Mode
Safety alarm messages on LCD
Skin Mode
- “Skin Temp. High” (>0.5°C of set temp.)
- “Skin Temp. Low” (<0.5°C of set temp.)
- “Skin Temp. Over Range” (>38°C)
- “Skin Temp. Over Range” (<34°C)
- “Skin Probe Fail”
Air Mode
- “Air Temp. High” (>1.5°C of set temp.)
- “Air Temp. Low” (<3.0°C of set temp.)
- “Air Temp. Over Range” (>39°C)
- “Air Probe Fail”
Safety high temp. cutoff at
- Skin temp. >38°C
- Air temp. >39°C
- Power Fail
- Heater Fail (optional)
Skin / Manual
Display
- 1” bright skin temp. display
- Set temp. display
- Numeric % of heater output display
- Selection of Skin & Manual modes
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Skin Mode
Safety alarm messages on LCD
- “Temp. High” (<0.5°C of set temp.)
- “Temp. Low” (>0.5°C of set temp.)
- “Temp. Over Range” (>38°C) /
- “Temp. Under Range” (<34°C)
- (Actual temp. starts blinking)
- “Probe Fail”
- Power failure
Safety high temp. cutoff at 38°C
For air control low temp. limit (<34°C) can be programmed.
INNOVA (NPRHWINV-01)- FIXED BASSINET + DOUBLESURFACE PHOTOTHERAPY
INNOVA (NPRHWINV-02)- FIXED BASSINET + UNDERSURFACE PHOTOTHERAPY
Components
Micro Computer Temperature Controller
- PID (proportionate integrated derivative) based micro processor technology assures accurate maintenance of set temperature with internal accuracy of 0.01°C
- User friendly on-board temperature calibration with feather touch keys
- User friendly feather touch control
- Key-pad lock facility to avoid accidental changes of any set parameters
- Safety alarm with high temperature cutoff
- Watch Dog facility Auto-Resets the System during hangups due to power fluctuation
Heater Box
- Ceramic Infra Red Heater with higher power output and more durability than any tubular silicon quartz glass heaters
- Swivelling source box to take X-ray
- Temperature Resistant Design of heater box keeps exterior surface cool even after continuous use
- Stainless steel reflector and heater grill for complete cleaning of heater & reflector without any electrical shock hazards
LED Examination Lamp
- Appropriate illumination to assess infant colour during procedures
Baby Bassinet (Stainless Steel)
- Easy access of an Infant from all sides due to clear acrylic collapsible side panels
- Acrylic Baby Tray for infant’s head Up / Down facility
- Provision for positioning X-Ray cassette
- Thick cushion mattress
- Nylon based moulded side pillars to fix the drop down sides
Essential Attachments
- Instrument Tray, Pole for Infusion Pump & IV Stand
Available Options
- Fabrication in Mild Steel or Stainless Steel
- Air/Pre-warm, Skin/Manual or Skin/Air/Manual (LCD) Controller
Temperature Controllers
Skin / Air / Manual (LCD)
Displays
- 1” bright skin temp. display
- 1” bright air temp. display
- Numeric % of heater output display
- Set temp. display on LCD
- Control mode display on LCD
- APGAR Timer display on LCD
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Auto Mode
Safety alarm messages on LCD
Skin Mode
- “Skin Temp. High” (>0.5°C of set temp.)
- “Skin Temp. Low” (<0.5°C of set temp.)
- “Skin Temp. Over Range” (>38°C)
- “Skin Temp. Over Range” (<34°C)
- “Skin Probe Fail”
Air Mode
- “Air Temp. High” (>1.5°C of set temp.)
- “Air Temp. Low” (<3.0°C of set temp.)
- “Air Temp. Over Range” (>39°C)
- “Air Probe Fail”
Safety high temp. cutoff at
- Skin temp. >38°C
- Air temp. >39°C
- Power Fail
- Heater Fail (optional)
Skin / Manual
Display
- 1” bright skin temp. display
- Set temp. display
- Numeric % of heater output display
- Selection of Skin & Manual modes
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Skin Mode
Safety alarm messages on LCD
- “Temp. High” (<0.5°C of set temp.)
- “Temp. Low” (>0.5°C of set temp.)
- “Temp. Over Range” (>38°C) /
- “Temp. Under Range” (<34°C)
- (Actual temp. starts blinking)
- “Probe Fail”
- Power failure
Safety high temp. cutoff at 38°C
For air control low temp. limit (<34°C) can be programmed.
INNOVA (NPRHWINV-02)- FIXED BASSINET + UNDERSURFACE PHOTOTHERAPY
INNOVA (NPRHWINV-03)- FIXED BASSINET + DRAWERS
Components
Micro Computer Temperature Controller
- PID (proportionate integrated derivative) based micro processor technology assures accurate maintenance of set temperature with internal accuracy of 0.01°C
- User friendly on-board temperature calibration with feather touch keys
- User friendly feather touch control
- Key-pad lock facility to avoid accidental changes of any set parameters
- Safety alarm with high temperature cutoff
- Watch Dog facility Auto-Resets the System during hangups due to power fluctuation
Heater Box
- Ceramic Infra Red Heater with higher power output and more durability than any tubular silicon quartz glass heaters
- Swivelling source box to take X-ray
- Temperature Resistant Design of heater box keeps exterior surface cool even after continuous use
- Stainless steel reflector and heater grill for complete cleaning of heater & reflector without any electrical shock hazards
LED Examination Lamp
- Appropriate illumination to assess infant colour during procedures
Baby Bassinet (Stainless Steel)
- Easy access of an Infant from all sides due to clear acrylic collapsible side panels
- Acrylic Baby Tray for infant’s head Up / Down facility
- Provision for positioning X-Ray cassette
- Thick cushion mattress
- Nylon based moulded side pillars to fix the drop down sides
Essential Attachments
- Instrument Tray, Pole for Infusion Pump & IV Stand
Available Options
- Fabrication in Mild Steel or Stainless Steel
- Air/Pre-warm, Skin/Manual or Skin/Air/Manual (LCD) Controller
Temperature Controllers
Skin / Air / Manual (LCD)
Displays
- 1” bright skin temp. display
- 1” bright air temp. display
- Numeric % of heater output display
- Set temp. display on LCD
- Control mode display on LCD
- APGAR Timer display on LCD
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Auto Mode
Safety alarm messages on LCD
Skin Mode
- “Skin Temp. High” (>0.5°C of set temp.)
- “Skin Temp. Low” (<0.5°C of set temp.)
- “Skin Temp. Over Range” (>38°C)
- “Skin Temp. Over Range” (<34°C)
- “Skin Probe Fail”
Air Mode
- “Air Temp. High” (>1.5°C of set temp.)
- “Air Temp. Low” (<3.0°C of set temp.)
- “Air Temp. Over Range” (>39°C)
- “Air Probe Fail”
Safety high temp. cutoff at
- Skin temp. >38°C
- Air temp. >39°C
- Power Fail
- Heater Fail (optional)
Skin / Manual
Display
- 1” bright skin temp. display
- Set temp. display
- Numeric % of heater output display
- Selection of Skin & Manual modes
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Skin Mode
Safety alarm messages on LCD
- “Temp. High” (<0.5°C of set temp.)
- “Temp. Low” (>0.5°C of set temp.)
- “Temp. Over Range” (>38°C) /
- “Temp. Under Range” (<34°C)
- (Actual temp. starts blinking)
- “Probe Fail”
- Power failure
Safety high temp. cutoff at 38°C
For air control low temp. limit (<34°C) can be programmed.
INNOVA (NPRHWINV-03)- FIXED BASSINET + DRAWERS
INNOVA (NPRHWINV-04)- FIXED BASSINET
Components
Micro Computer Temperature Controller
- PID (proportionate integrated derivative) based micro processor technology assures accurate maintenance of set temperature with internal accuracy of 0.01°C
- User friendly on-board temperature calibration with feather touch keys
- User friendly feather touch control
- Key-pad lock facility to avoid accidental changes of any set parameters
- Safety alarm with high temperature cutoff
- Watch Dog facility Auto-Resets the System during hangups due to power fluctuation
Heater Box
- Ceramic Infra Red Heater with higher power output and more durability than any tubular silicon quartz glass heaters
- Swivelling source box to take X-ray
- Temperature Resistant Design of heater box keeps exterior surface cool even after continuous use
- Stainless steel reflector and heater grill for complete cleaning of heater & reflector without any electrical shock hazards
LED Examination Lamp
- Appropriate illumination to assess infant colour during procedures
Baby Bassinet (Stainless Steel)
- Easy access of an Infant from all sides due to clear acrylic collapsible side panels
- Acrylic Baby Tray for infant’s head Up / Down facility
- Provision for positioning X-Ray cassette
- Thick cushion mattress
- Nylon based moulded side pillars to fix the drop down sides
Essential Attachments
- Instrument Tray, Pole for Infusion Pump & IV Stand
Available Options
- Fabrication in Mild Steel or Stainless Steel
- Air/Pre-warm, Skin/Manual or Skin/Air/Manual (LCD) Controller
Temperature Controllers
Skin / Air / Manual (LCD)
Displays
- 1” bright skin temp. display
- 1” bright air temp. display
- Numeric % of heater output display
- Set temp. display on LCD
- Control mode display on LCD
- APGAR Timer display on LCD
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Auto Mode
Safety alarm messages on LCD
Skin Mode
- “Skin Temp. High” (>0.5°C of set temp.)
- “Skin Temp. Low” (<0.5°C of set temp.)
- “Skin Temp. Over Range” (>38°C)
- “Skin Temp. Over Range” (<34°C)
- “Skin Probe Fail”
Air Mode
- “Air Temp. High” (>1.5°C of set temp.)
- “Air Temp. Low” (<3.0°C of set temp.)
- “Air Temp. Over Range” (>39°C)
- “Air Probe Fail”
Safety high temp. cutoff at
- Skin temp. >38°C
- Air temp. >39°C
- Power Fail
- Heater Fail (optional)
Skin / Manual
Display
- 1” bright skin temp. display
- Set temp. display
- Numeric % of heater output display
- Selection of Skin & Manual modes
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Skin Mode
Safety alarm messages on LCD
- “Temp. High” (<0.5°C of set temp.)
- “Temp. Low” (>0.5°C of set temp.)
- “Temp. Over Range” (>38°C) /
- “Temp. Under Range” (<34°C)
- (Actual temp. starts blinking)
- “Probe Fail”
- Power failure
Safety high temp. cutoff at 38°C
For air control low temp. limit (<34°C) can be programmed.
INNOVA (NPRHWINV-04)- FIXED BASSINET
INNOVA (NPRHWLT-01) - FIXED BASSINET LITE
Components
Micro Computer Temperature Controller
- PID (proportionate integrated derivative) based micro processor technology assures accurate maintenance of set temperature with internal accuracy of 0.01°C
- User friendly on-board temperature calibration with feather touch keys
- User friendly feather touch control
- Key-pad lock facility to avoid accidental changes of any set parameters
- Safety alarm with high temperature cutoff
- Watch Dog facility Auto-Resets the System during hangups due to power fluctuation
Heater Box
- Ceramic Infra Red Heater with higher power output and more durability than any tubular silicon quartz glass heaters
- Swivelling source box to take X-ray
- Temperature Resistant Design of heater box keeps exterior surface cool even after continuous use
- Stainless steel reflector and heater grill for complete cleaning of heater & reflector without any electrical shock hazards
LED Examination Lamp
- Easy access of an Infant from all sides due to clear acrylic collapsible side panels
- Bed head Up / Down facility
- Provision for positioning X-Ray cassette
- Thick cushion mattress
- Basinet side cover easy to be open & close without difficulty
- Self locking single hand operation
- The side rails foldable to care the baby and transparent basinet tilting facility
Baby Bassinet (Stainless Steel)
- Easy access of an Infant from all sides due to clear acrylic collapsible side panels
- Acrylic Baby Tray for infant’s head Up / Down facility
- Provision for positioning X-Ray cassette
- Thick cushion mattress
- Nylon based moulded side pillars to fix the drop down sides
Essential Attachments
- Instrument Tray, Pole for Infusion Pump & IV Stand
- MCB (Optional)
- Manual Height adjustment (Optional)
Skin / Manual Temperature Controller
Displays
- 1” bright skin temp. display
- Set temp. display
- Numeric % of heater output display
- Selection of Skin & Manual modes
Manual Mode
- Time programmable
- Heater output adjustable
- Over temp. cutoff.
- Timer / APGAR Timer
- Audio Alarm mute facility
- Mute time programmable
Skin Mode
Safety alarm messages on LCD
- “Temp. High” (<0.5°C of set temp.)
- “Temp. Low” (>0.5°C of set temp.)
- “Temp. Over Range” (>38°C) / “Temp. Under Range” (<34°C) (Actual temp. starts blinking)
- “Probe Fail”
- Power failure
Safety high temp. cutoff at 38°C
- Skin Temp. – 28-37°C
- Display Temp. – 10-50°C
- APGAR Timer
- Working Temp.: 10-40°C
- Relative Humidity: 25-75%
- Atmospheric Pressure: 500 – 900 mmHg
- AC – 220 VAC – 50 Hz
- 4 Castor wheels, 2 lockable
Tests
Tested as per following IEC Standards :
IEC 601-1:1998 + A1:1991 + A2:1995
IEC 601-2-21:1994
IEC 60601-1-2, 2001, 2nd Edition
Amendment A1:1996 to IEC 60601-2-21:1994
INNOVA (NPRHWLT-01) - FIXED BASSINET LITE
INNOVA (NPRHWINV) - STAND
Components
Micro Computer Temperature Controller
- PID (proportionate integrated derivative) based micro processor technology assures accurate maintenance of set temperature with internal accuracy of 0.01°C
- User friendly on-board temperature calibration with feather touch keys
- User friendly feather touch control
- Key-pad lock facility to avoid accidental changes of any set parameters
- Safety alarm with high temperature cutoff
- Watch Dog facility Auto-Resets the System during hangups due to power fluctuation
Heater Box
- Ceramic Infra Red Heater with higher power output and more durability than any tubular silicon quartz glass heaters
- Swivelling source box to take X-ray
- Temperature Resistant Design of heater box keeps exterior surface cool even after continuous use
- Stainless steel reflector and heater grill for complete cleaning of heater & reflector without any electrical shock hazards
LED Examination Lamp
- Appropriate illumination to assess infant colour during procedures
Baby Bassinet (Stainless Steel)
- Easy access of an Infant from all sides due to clear acrylic collapsible side panels
- Acrylic Baby Tray for infant’s head Up / Down facility
- Provision for positioning X-Ray cassette
- Thick cushion mattress
- Nylon based moulded side pillars to fix the drop down sides
Essential Attachments
- Instrument Tray, Pole for Infusion Pump & IV Stand
Available Options
- Fabrication in Mild Steel or Stainless Steel
- Air/Pre-warm, Skin/Manual or Skin/Air/Manual (LCD) Controller
Temperature Controllers
Skin / Air / Manual (LCD)
Displays
- 1” bright skin temp. display
- 1” bright air temp. display
- Numeric % of heater output display
- Set temp. display on LCD
- Control mode display on LCD
- APGAR Timer display on LCD
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Auto Mode
Safety alarm messages on LCD
Skin Mode
- “Skin Temp. High” (>0.5°C of set temp.)
- “Skin Temp. Low” (<0.5°C of set temp.)
- “Skin Temp. Over Range” (>38°C)
- “Skin Temp. Over Range” (<34°C)
- “Skin Probe Fail”
Air Mode
- “Air Temp. High” (>1.5°C of set temp.)
- “Air Temp. Low” (<3.0°C of set temp.)
- “Air Temp. Over Range” (>39°C)
- “Air Probe Fail”
Safety high temp. cutoff at
- Skin temp. >38°C
- Air temp. >39°C
- Power Fail
- Heater Fail (optional)
Skin / Manual
Display
- 1” bright skin temp. display
- Set temp. display
- Numeric % of heater output display
- Selection of Skin & Manual modes
Manual Mode
- Time programmable
- Heater output adjustable
- Over temperature cutoff
- Timer / APGAR Timer
- Baby attending alarm
- Mute time programmable
- Audio alarm with mute facility
Skin Mode
Safety alarm messages on LCD
- “Temp. High” (<0.5°C of set temp.)
- “Temp. Low” (>0.5°C of set temp.)
- “Temp. Over Range” (>38°C) /
- “Temp. Under Range” (<34°C)
- (Actual temp. starts blinking)
- “Probe Fail”
- Power failure
Safety high temp. cutoff at 38°C
For air control low temp. limit (<34°C) can be programmed.